Frequently asked questions

How Lumero measures particle size, what it adds to DLS, and how its results are validated. If your question isn't answered here, ask us directly.

How does Lumero® work?

Lumero records a microscopy video of your sample and analyses the intensity fluctuations within it to calculate particle size. The method is called VISTA, or Video Intensity Spatio-Temporal Analysis.

A single Lumero measurement can contain billions of nanoscale particles. At any moment there are fluctuations in particle density caused by their random position and motion. VISTA extracts those fluctuations and analyses them much as dynamic light scattering (DLS) does, obtaining diffusion coefficients and from those the hydrodynamic radii of the particles. The principles and the equipment are very different, but a Lumero measurement can be thought of as combining the strengths of DLS and microscopy while avoiding what limits each of them: the dilution DLS requires, and the resolution limit of the microscope. Because the analysis works on intensity fluctuations rather than on individual particles, it resolves particles below that limit.

How do I know I can trust Lumero?

Every Lumero measurement keeps the microscopy video it came from, so the large particles typically missed by other techniques are visible to the eye rather than inferred.

Lumero is built on more than 50 years of combined research experience in Dyneval’s technical team, and has been tested on a wide range of reference samples and against a wide range of established techniques. Recent work by the National Physical Laboratory compared Lumero against dynamic light scattering (DLS), particle tracking analysis (PTA), small-angle X-ray scattering (SAXS), differential centrifugal sedimentation (DCS) and laser diffraction (LD). It found excellent agreement across those techniques, and the widest range of accessible length scales in a single instrument.

What does Lumero add to DLS?

DLS measures dilute, monodisperse suspensions accurately. Lumero does the same for concentrated and complex suspensions, so you see the sample in the conditions that matter rather than in a diluted proxy of it.

Removing the dilution step removes its artefacts: dilution can disperse aggregates or create them, it changes the aggregation state, and every handling step is an opportunity for error. It also allows changes to be followed in real time, with around five minutes of setup.

What is a "complete" particle size distribution?

A complete particle size distribution resolves every population in a sample at once, across a wide size range, rather than measuring only the range a given technique is biased towards.

Most techniques size monodisperse particles well across a wide range. When several populations are present, that usually stops being true. DLS is often assumed to be biased towards large particles, but in practice the form scattering factor conceals large particles that Lumero detects. Those particles may be a few per cent of the total mass while being only one in a million by number, so in a single-particle technique such as PTA, which might count a thousand particles, they are missed entirely. Lumero reports them in the distribution, and the microscopy video confirms they are real.

How much sample does Lumero require?

Between 3 and 30 µL, depending on the sample chamber. The standard measurement uses 20 µL.

Because Lumero often measures the sample undiluted, users want to use as little as possible, particularly where components are expensive. Chambers are available that support measurement with as little as 3 µL.

What applications is Lumero designed for?

Formulation development, stability studies, filtration optimisation and quality control.

Whether you are monitoring particle size, investigating aggregation, evaluating freeze-thaw stability or validating product quality, Lumero gives rapid, unbiased particle data to support the decision in front of you.

Can Lumero measure powders?

Not directly. Lumero characterises particles dispersed in a liquid.

Powders can usually be analysed once dispersed in a suitable liquid, which often represents the final application or formulation better than a dry measurement would. For guidance on a specific powder, our team can advise on the most appropriate approach.